Question & Responses for Buy # 1183653

Questions & Responses Released to All Sellers
Buy #: 1183653
Question Deadline:  None
Question Count: 2
Buy #: 1183653
Question Deadline:  None
Question Count: 2
Question Submitted:
Can you please explain what is needed for the "Authorized Manufacturer Rep" form being requested?
Response Received:
Proof of authorization can consist of: Cover letter of franchise agreement with the manufacturer. Letter on manufacturer's letterhead or an email from the manufacturer's official web domain stating that the distributor is authorized/franchised or Proof of listing as authorized/franchised on manufacturer's web site.
Question Submitted:
1) The ELSD detector needs a nitrogen supply, but I didn’t see that called out in the bid. a. This is what it says: Detectors: Evaporative Light Scattering Detector (ELSD): Input gas: high purity nitrogen gas or zero grade air b. Can you confirm if that is needed? We work with PEAK for high purity gas supply. 2) How do you imagine the detectors set up? Depending on how you would want the flowpath to be, you might need a splitter to divert the flow-path to different detectors. If you had the flowpath as RI -> PDA -> ELSD, then you wouldn’t need the splitter.
Response Received:
A1: - We are requesting that the ELSD input gas be compatible with either high purity nitrogen and/or zero grade air. No generator is necessary for this. The ELSD just needs the appropriate connections for a gas cylinder. A2: - No splitter is needed.
Unison Marketplace Process Help
Help and Training page which includes training videos, interactive help, and more. A link can also be found on the top right of every page, after log in.

For additional information, please contact:

Marketplace Support
Tel: (877) 933-3243
Contact Marketplace Support